![background image](http://html.mh-extra.com/html/zeiss/crossbeam-340/crossbeam-340_instruction-manual_927492055.webp)
PK=aÉëÅêáéíáçå
fåëíêìÅíáçå=j~åì~ä=`êçëëÄÉ~ã=PQM
=
ÉåMOE`loF
=RR=çÑ=NPQ
qÉÅÜåáÅ~ä=Ç~í~
3.5.4. System layout
1
2
1
1
3
4
5
2
6
3
4
5
No
Description
Size (mm)
approx.
Weight (kg)
approx.
Distribution
of load (kg)
Footprints
1
column
940 x 1083 x 1750
870
4 x 222.5
4 x Ø 80 mm
Capella FIB
940 x 1082 x 1700
920
4 x 230
4 x Ø 80 mm
Capella FIB
+ 200mm-airlock
1060 x 1082 x 1700
950
4 x 238
4 x Ø 80 mm
200mm-air-
lock
940 x 1082 x 1700
900
4 x 225
4 x Ø 80 mm
2
Table + PC
1153 x 980 x 1350
97
4 x 24.3
4 x Ø 50 mm
3
Dynamic damping block
200 x 200 x 290
26
1 x 26.0
200 mm x 200 mm
4
Static damping block
180 x 180 x 160
36.6
1 x 36.6
180 mm x 180 mm
5
Pre-vacuum pump
430 x 250 x 290
24.5
1 x 24.5
200 mm x 180 mm