UK=qêçìÄäÉëÜççíáåÖ
fåëíêìÅíáçå=j~åì~ä=`êçëëÄÉ~ã=PQM
=
ÉåMOE`loF
=NNT=çÑ=NPQ
`çäìãå
The
FIB Gun Presssure
must be better than 5
x 10
-7
mbar.
CAUTION
Risk of arcing. Danger of damaging the ion source.
Before switching on the ion beam, ensure the FIB gun pressure is better than 5 x 10
-7
mbar.
2
To switch on the ion gun, click
ON
.
CAUTION
Risk of damaging the ion source.
Ensure the FIB gun is switched ON before you start the heating procedure.
3
Click
Heating
.