6.11. Changing to STANDBY mode
Change to
STANDBY
mode if the microscope is not operated for longer periods, for example over
the weekend.
The filament will continue to be heated, and the vacuum in electron optical column and specimen
chamber will be maintained.
Procedure
1
Switch off the EHT, see section 6.10.
2
Log off from SmartSEM
®
3
Close the EM Server.
A window appears asking for confirmation to
close the session.
4
Click
Yes
.
5
Shut down the PC.
6
At the plinth of the microscope, press the
STANDBY
button.
The yellow
STANDBY
push button is blinking,
then permanently on.
We also recommend the
STANDBY
mode to store the microscope.
In this case, select the
Partial Vent on Standby
checkbox in the
Vacuum
tab of the
SEM Control
panel. This maintains the gun vacuum, and switches off and protects the turbo pump.