lÄí~áåáåÖ=~=Ñáêëí=áã~ÖÉ
6.4.4. Switching on the gun
If the gun is switched on, go to section 6.4.5.
1
In the
Vacuum
tab, check that
EHT Vac ready=Yes
is indicated.
2
To open the pop-up menu, in the status bar,
click
Gun
.
3
In the pop-up menu, select
Gun On
.
The gun runs up.
This may take up to 5 minutes.
CAUTION
Risk of damaging the Schottky field emitter.
Avoid switching off the gun during the working week to optimise the life time of the Schott-
ky field emitter.
Switching off the gun is only recommended at longer breaks of two days or more, such as
weekends.