If it is necessary to install filters without direction
arrows, we recommend the following procedure:
Filters with reflective, dielectric layers need to be
installed, so that the reflective layer (2-43/
6
) on
the excitation filter (2-43/
5
) points outwards (in
relation to the reflector module). With the
emission filter (2-43/
1
), the reflective layer (2-43/
2
)
points inwards (Fig. 2-43).
The reflective layer (2-43/
4
) of the beam splitter
(2-43/
3
) should point downwards in its installation
position.
The arrows (2-43/
7
) mark the illumination or
imaging beam path.
Fig. 2-43
Installing the filter and beam