Axioskop 40
START-UP
Axioskop 40 FL
Initial start-up
Carl Zeiss
B 40-810 e 12/01
2-17
2.1.13
HBO 50 fluorescence illuminator
(Axioskop 40 FL)
Attaching the HBO 50 fluorescence
illuminator
•
Place the HBO 50 fluorescence illuminator
(2-21/
3
) on the connection tube (2-21/
1
) on the
Axioskop 40 FL stand, align and secure it with
clamping screw (2-21/
2
).
•
Connect the HBO 50 connection cable to the
transformer and the power supply cable to a
power outlet.
•
Turn the HBO 50 fluorescence illuminator on
and off with the power switch on the front of
the transformer. The lamp will ignite
automatically when switched on.
Adjusting the fluorescence illuminator
Providing the Axioskop 40 FL is fitted with an
adjustment guide (to be ordered separately), the
HBO 50 can remain attached to the epi-
fluorescence illuminator while the burner is
adjusted.
•
Switch on the completely connected
fluorescence illuminator via the transformer.
The burner ignites automatically.
If the arc image and reflection are
superimposed, the burner will have
a higher thermal load, resulting in a
shorter service life.
•
Pull out adjustment guide (2-22/
5
). The arc of
the burner (lighter) and its reflection (slightly
darker) are displayed in the adjustment guide
viewing window (2-22/
6
).
Fig. 2-21
Attaching HBO 50
Fig. 2-22
Adjusting HBO 50
Summary of Contents for Axioskop 40
Page 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Page 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Page 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Page 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Page 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Page 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...