![background image](http://html1.mh-extra.com/html/zeiss/axioskop-40/axioskop-40_operating-manual_3382221036.webp)
START-UP
Axioskop 40
Carl Zeiss
Initial start-up
Axioskop 40 FL
2-6
B 40-810 e 12/01
Inserting the eyepiece reticule
The PL 10x/23 Br. foc. eyepieces are intended for
use with eyepiece reticules.
The slight image shift caused by the additional
path through glass is taken into account on the
diopter scale by the fact that the zero point
position is indicated not by the white dot (2-4/
W
),
but by the red dot (2-4/
R
).
The eyepiece reticules (2-4/
3
) have been adhered
to screw-in mounts (2-4/
4
) by the manufacturer
for easy replacement. Complete mounts with
reticules adhered can be ordered directly from
Zeiss.
To change the mount, proceed as follows:
•
To replace an eyepiece reticule, unscrew the
mount (2-4/
4
) with eyepiece reticule (2-4/
3
)
from the eyepiece. Replace it with a new
mount containing the eyepiece reticule
required.
☞
If you insert eyepiece reticules into the unscrewed mount, take care to ensure that the labeling
is visible the right way up after insertion.
Compensation of ametropia when eyepiece reticules are used
Correct use of an eyepiece reticule requires two focusing eyepieces, e.g. PL 10x/23 Br. foc., to enable the
user to compensate for differences in the visual acuity of his or her eyes.
•
Use the focusing lens of the adjustable eyepiece to focus on the line figure of the eyepiece reticule.
•
Use the focusing drive to focus on the microscope image of a specimen through the eyepiece with
reticule.
•
As soon as the microscope image and eyepiece reticule are focused, use the second focusing eyepiece
to focus the microscope image for the second eye.
The two microscope images and reticules will now be focused.
From this point in time on, you should focus only with the focusing drive.
Fig. 2-4
Inserting the eyepiece reticule
Summary of Contents for Axioskop 40
Page 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Page 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Page 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Page 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Page 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Page 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...