START-UP
Axioskop 40
Carl Zeiss
Attachments and conversions
Axioskop 40 FL
2-40
B 40-810 e 12/01
2.2.15
Attaching the active co-observation equipment for 2 users
The active co-observation equipment for 2 users can be attached to the Axioskop 40 stand without
restrictions. For the Axioskop 40 FL stand, the incident-light illuminator (e.g. HBO 103) must first be
removed.
•
Use a SW 3 ball-headed screwdriver to undo clamping screw (2-47/
12
) and remove available tube
upwards.
•
Remove dust cap (2-47/
14
) from the dovetail at (2-47/
15
) of the co-observation equipment (2-47/
16
).
•
Insert the dovetail of the co-observation equipment in the stand opening (2-47/
13
) and tighten
clamping screw (2-47/
12
) only slightly at first.
•
Align the co-observation equipment with the outer edges of the stand and ensure that the support
points on the co-observation equipment (bottom) and the stand (2-47/
9
) match exactly.
•
Then tighten clamping screw (2-47/
12
) to prevent the co-observation equipment from shifting.
•
Unscrew the tube lenses from the two binocular tubes (using the cover of one of the storage cases)
and put them into the cases.
•
Remove upper dust caps (2-47/
3
and
18
) from the co-observation equipment.
•
Place the ergonomic binocular phototube 6-25°/23 (100/100) (2-47/
1
) onto the front dovetail
(2-47/
17
) of the co-observation equipment. In addition to the ergonomic binocular phototube, the
binocular tube 30°/23 (2-47/
2
) can be placed onto the rear dovetail (2-47/
4
). Then align the tubes and
tighten with clamping screws (2-47/
8
and
19
).
The ergonomic binocular phototube 6-25°/23 provide an upright and unreversed image of the specimen
in the eyepiece.
•
Screw heat protection shield (2-47/
6
) to co-observation equipment from below using the two
fastening screws (2-47/
7
).
•
Insert power unit (2-47/
10
, 115 V or 230 V) plug (2-47/
11
) into the socket (2-47/
5
) on the co-
observation equipment and tighten the ring to prevent shifting.
•
Connect power unit to a power outlet (115 V or 230 V).
Summary of Contents for Axioskop 40
Page 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Page 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Page 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Page 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Page 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Page 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...