START-UP
Axioskop 40
Carl Zeiss
Attachments and conversions
Axioskop 40 FL
2-30
B 40-810 e 12/01
•
Press the stage to the front against the spring pin and lower towards the back of stage carrier
(2-35/
3
), then release it.
•
Tighten screw cap (2-35/
2
).
c)
Centering the stage
When objectives with high magnification are used,
centering can be exact for one selected objective
only.
All stages are factory-precentered, i.e. a set
specimen detail remains in the image center when
the stage is rotated. If the image detail moves
from the image center (2-36/
5
) with the stage
rotating, recentering should be performed as
follows:
•
Undo stage clamping screw (2-36/
4
) and the
screw cap on the stage carrier (2-36/
1
).
•
Rotate the stage to determine the maximum
specimen deflection (2-36/5, tip of the arrow)
in the direction of the eyepiece reticule.
•
Reset the two centering screws on the stage
carrier (2-36/
2
) using one SW 1.5 Allen
screwdriver (2-36/
3
) each to move the
specimen detail by half the arrow length in the
direction of the crossline center. Check whether
the specimen detail moves when the stage is
rotated again; repeat the procedure, when
required.
•
When centering is finished, tighten screw cap
(2-36/
1
) again.
The stage can be turned through 240° up to a
range of y
≤
27 mm. No rotation is possible above
this range of movement.
Fig. 2-36
Centering the rotary mechanical
stage
Summary of Contents for Axioskop 40
Page 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Page 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Page 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Page 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Page 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Page 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...