Axioskop 40
START-UP
Axioskop 40 FL
Attachments and conversions
Carl Zeiss
B 40-810 e 12/01
2-35
2.2.12
Installing and removing "Push&Click" modules on the Axioskop 40 FL
The FL P&C reflector modules and optovar and analyzer modules can only be installed on the Axioskop
40 FL, since its stand is equipped with a reflector turret. The following reflector modules can be installed:
FL P&C reflector module
# 000000-1046-281
Optovar module 1.25x (for transmitted light only)
# 000000-1046-284
Optovar module 1.6x (for transmitted light only)
# 000000-1046-283
Optovar module 2.5x (for transmitted light only)
# 000000-1046-282
Analyzer module D
# 000000-1050-958
Installing a module:
•
Pull the cover cap (2-41/
1
) forwards off the
reflector turret.
•
Insert the module (2-41/
3
) with the retaining
pins (2-41/
4
) from above to the left and right of
the module at an angle into the spring clamps
(2-41/
2
) on the reflector turret.
•
Then press the module at the top until it snaps
securely into the top spring clamps of the
reflector turret.
Aligning the module when installing it in the
reflector turret:
FL P&C reflector module
the excitation filter faces forward; the emission
filter faces up
Optovar module
the lens faces up; the empty module opening
faces forward
Analyzer module
the analyzer faces up; the empty module
opening faces forward
Fig. 2-41
Changing the FL P&C reflector
module
Summary of Contents for Axioskop 40
Page 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Page 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Page 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Page 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Page 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Page 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...