Axioskop 40
START-UP
Axioskop 40 FL
Attachments and conversions
Carl Zeiss
B 40-810 e 12/01
2-27
Avoid fingerprints on the burner’s glass parts. Where necessary, remove same immediately.
•
Press down spring lever (2-32/
4
) and insert the burner with the heat sink into lamp holder (2-32/
5
).
Take care to ensure you only hold the heat sink.
•
Slowly let go of the spring lever and release the heat sink.
The heat sink must be aligned parallel to the lamp base. To align it, press the spring lever and
turn the heat sink with the burner in the lamp holder.
•
Replace the lamp housing and tighten clamping screw (2-32/
1
). Log the number on the operational
hour counter.
The burner must be replaced after it attains its service life of 100 hours.