Axioskop 40
OPERATION
Axioskop 40 FL
Axioskop 40 FL operation and function controls
Carl Zeiss
B 40-810 e 12/01
3-15
Luminous-field diaphragm (3-7/22)
See page 3-7.
Drive knob for condenser height adjustment
(3-7/23)
−
Height adjustment of the condenser to set
KÖHLER illumination.
Clamping lever for focusing drive height stop
(3-7/24)
−
Adjustable height stop for stage focusing..
•
Turn clamping lever up against the pin stop.
Move stage to the highest position with the
focusing drive to prevent collisions with the
stage or carrier plate. Clamp in position by
pressing on the clamping lever.
Fig. 3-11
Luminous-field diaphragm