Chapter
8
■
About inputs
■
About the RUN signal
The RUN output stops when an abnormal state of the indicated value is detected. The conditions
under which the output stops as in an abnormal state are as follows.
- ±LOAD
- OFL1, OFL2
- Calibration error
- Digital zero limit error
* See "12-7.Error/message list " on page 131 about the contents of errors.
8-2.
SI/F (serial interface)
■
SI/F indicated value format
GROSS area:
Torque
NET area:
Rotation speed
TARE area:
Angle
■
SI/F print
《
Setting value
》
0: None
No print command is sent.
1: Stable value
When stabilityis detected, a print command is sent to the SI/F to make the
connected printer print automatically. When near zero is OFF, the torque
is held until near zero ON after stable ON.
Hold is reset in 1.5 sec. after near zero ON.
2: Hold value
When hold is reset, a print command is sent to make the connected
printer print the hold value automatically.
(When the hold mode is OFF, turn on T/H or press
.
)
B7
Common for external input signals.
→
"
■
External I/O connection " on page 14
B8, B9
Inputs the hold control signal.
→
"7.EXPLANATION OF HOLD
FUNCTIONS " on page 55
B10
Inputs the zero clear (Angle) signal.
→
"5-7.Zero clear (Angle) " on page 44
B11
Inputs the Digital Zero signal.
→
"5-6.Digital zero " on page 43
B12
Input for record data clear.
→
"9-3.Record data clear " on page 78
OFF
RUN ON
OFF
RUN
ON
0.5sec. pulse
3.0sec
Error factor removal
This signal turns off when an error occurs.
duty50%
* By the error factor removal, ON-OFF operation restarts in 3.0 sec.
* RUN turns off when the power is turned off.