71
8
STANDARD INTERFACE
71
S
T
ANDARD INT
E
RFACE
Chapter
8
8-3.
RS-485
Usable when connecting to UTM
Ⅲ
& UTF (TM380 option).
You can use RS-485 to change settings, digital zero, display torque and rotation speed of UTM
Ⅲ
&
UTF (TM380option).
Communication specifications
Signal level
Based on RS-485
Transmitting distance
Approx. 1 km
Transmitting method
Asynchronous, Full duplex
Transmitting speed
230.4kbps
Bit configuration
Start bit: 1bit
Character length: 8bit
Stop bit: 1bit
Parity: none
Terminator: CR
Code
ASCII
■
RS-485 communication mode
In communication mode, select RS-485 command mode / RS-485 display mode / invalid.
《
Setting value
》
0: RS-485 command mode 1: RS-485 display mode 2: Invalid
1.
Select setting mode 6.
Press
→
five times.
2.
Set "RS-485 communication mode."
Press
once.
⇒
"RS-485 communication mode" is
displayed.
Press
, and present set value blinks.
Input the setting value with
.
Press
to validate it. (Initial value: 0)
3.
Press
twice to go back to the Indicated value display.
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
ESC