29
4
CALIBRATION PROCEDURE
29
CALIBRATION PROCE
DURE
Chapter
4
◇
Actual load calibration setting method
1.
Select setting mode 4.
Press
→
three times.
2.
Set "Actual load" of the sensor.
Press
four times.
⇒
"Actual load calibration" is
displayed.
Enter a setting value input state with
, and the
highest digit blinks.
Check the capacity of the actually-applied load, and input
the
value
with .
Move to the lower digits with
. (Initial value: 10000)
Recheck that the load is properly applied on the sub
display (angle display side), and press
to validate
it.
3.
Actual load calibration is executed.
If actual load calibration is executed normally, a message
as shown on the right-hand side is displayed.
If actual load calibration is not executed normally, the
alarm sounds and an error is displayed.
4.
If actual load calibration is executed normally, you
will go back to the setting item select state.
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
Torque
Speed
rpm
Nm
H I
LO
HOLD
H I
LO
HOLD
Angle
rad deg
Angle
rad deg
RECALL
RECALL
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
Summary of Contents for TM380
Page 1: ...TM380 20MAY2021REV 1 02 TORQUE MONITOR OPERATION MANUAL ...
Page 11: ...Contents X Contents X M E M O ...
Page 147: ......