45
5
EXPLANATION OF INDICATED- VALUE-RELATED FUNCTIONS
45
EXP
LANATION OF
INDICATED-
VALUE-RELATED FUNCTIONS
Chapter
5
Setting example
Motion detect (time)
0.4 sec.
Motion detect (range)
010
When the condition of dn
≦
set count (010) continues for the set time (0.4 sec.) or more
(d1
≦
10, and d2
≦
10, and ..., and d8
≦
10), the indicated value is regarded as stable.
■
Motion detect setting method
1.
Select setting mode2.
Press
→
once.
2.
Set "Motion detect (time)."
Press
once.
⇒
"Motion detect (time)" is displayed.
Press
, and the highest digit blinks.
Input the setting value with
.
Move to the lower digits with
. (Initial value: 0.0)
Press
to validate it.
3.
Set "Motion detect (range)."
Press
once.
⇒
"Motion detect (range)" is
displayed.
Press
, and the highest digit blinks.
Input the setting value with
.
Move to the lower digits with
. (Initial value: 000)
Press
to validate it.
4.
Press
twice to go back to the indicated value display.
d
1
d
2
d
3
d
4
d
5
d
6
d
7
d
8
Motion detect time (0.4sec.)
50msec.
dn = 50msec. difference of the previous
Indicated value
Time
instruction value
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
ESC