CALIBRATION PROCEDURE
25
CALIBRATION PROCE
DURE
Chapter
4
③
Unit (torque) setting
Set the unit for performing calibration.
After this setting, make sure to perform calibration.
≪
Setting value
≫
00: mNm 01: Ncm 02: Nm 03: kNm 04: kgm 05: kgcm 06: gcm
④
Decimal place (torque) setting
Set the decimal place for load-value-related display, setting items, etc.
≪
Setting value
≫
0: None 1: 0.0 2: 0.00 3: 0.000
⑤
Min. scale division (torque) setting
(This step may be omitted if there is no change.)
Set the minimum unit (scale interval, scale division) of the torque.
≪
Setting value
≫
0: 1 1: 2 2: 5 3: 10 4: 20 5: 50 6: 100
◇
How to set the unit, decimal place, and min. scale division of torque
1.
Select setting mode 4.
Press
→
three times.
2.
Set "Unit," "Decimal place," and "Min. scale division"
of torque.
Press
once.
⇒
"Calibration function select 1" is
displayed.
Press
, and the highest digit blinks.
①
Select unit.
Input the unit with
, and move to the
lower digit with
. (Initial value: 02)
②
Select decimal place.
Input the decimal place with
, and move
to the lower digit with
. (Initial value: 2)
③
Select min. scale division.
Input the min. scale division with
, and
validate it with
. (Initial value: 0).
3.
Press
twice to go back to the indicted value display.
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
0
2
2
0
《
Setting value
》
Min. scale division
Decimal place
0: None
1: 0.0
2: 0.00
3: 0.000
0: 1
1: 2
2: 5
3: 10
4: 20
5: 50
6: 100
Unit (torque)
00: mNm
01: Ncm
02: Nm
03: kNm
04: kgm (kgfm)
05: kgcm (kgfcm)
06: gcm (gfcm)
(torque)
(torque)
RECALL
RECALL
RECALL
RECALL
ESC