91
10
OPTION INTERFACE
91
OP
TION IN
TERFA
C
E
Chapter
10
3. Send/receive [operation command]
4. Send/receive [returning an error]
Movement directive
Function
Header
Protocol
Main
Sub
Zero
calibration
C
Z
Equivalent
input
calibration
C
S
Digital zero
C
G
Digital zero
reset
C
H
Record data
clear
C
J
Zero clear
(Angle)
C
K
SECTION
ON
C
1
SECTION
OFF
C
2
T/H ON
C
3
T/H OFF
C
4
Returning an error
Function
Header
Protocol
Main
Sub
Error
occurrence
(send/receive
mode)
E
R
C
CR
Host
TM380
Z
C
CR
Host
TM380
S
* Set the rated capacity value before sending an actual load calibration command.
C
CR
Host
TM380
G
C
CR
Host
TM380
H
C
CR
Host
TM380
J
C
CR
Host
TM380
K
C
CR
Host
TM380
1
C
CR
Host
TM380
2
C
CR
Host
TM380
3
C
CR
Host
TM380
4
If the calibration protect is ON, zero calibration and span calibration cannot be
performed.
Point
Command
*
R
E
Delimiter
Host
TM380
1: The command or message length is improper.
2: The protect of the setting value you are going to write is ON, or
a calibration command is received while the calibration protect is ON.
3: The format errors of the setting value which tried to write in.
4: Other communication errors (parity error, frame error, etc.)