37
4
CALIBRATION PROCEDURE
37
CALIBRATION PROCE
DURE
Chapter
4
④
Direction
Set the rotation direction of torque meter.
≪
Setting value
≫
0: NORMAL
Select this option when using the torque meter counterclockwise
viewing from the drive side.
1: REVERSE
Select this option when using the torque meter clockwise viewing
from the drive side.
◇
Direction setting method
1.
Select setting mode 7.
Press
→
six times.
2.
Set "Direction."
Press
four times.
⇒
"Direction" is displayed.
Press
, and present set value blinks.
Input the setting value with
.
(Initial value: 0)
Validate it with
.
3.
Press
twice to go back to the indicated value display.
⑤
Minimum scale (angle)
(This step may be omitted if there is no change.)
Set the minimum unit of rotary encoder such as scale interval and scale division.
≪
Setting value
≫
- When the unit (angle) is [0: deg]
0: 0.1 1: 0.2 2: 0.5 3: 1 4: 2 5: 5 6: 10 7: 20 8: 50 9: 100
- When the unit (angle) is [1: rad]
3: 0.01 4: 0.02 5: 0.05 6: 0.1 7: 0.2 8: 0.5 9: 1
◇
Minimum scale (angle) setting method
1.
Select setting mode 7.
Press
→
six times.
2.
Set "Minimum scale (angle)."
Press
five times.
⇒
"Minimum scale (angle)" is
displayed.
Press
, and present set value blinks.
Input the setting value with
.
(Initial value: 3)
Validate it with
.
3.
Press
twice to go back to the indicated value display.
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
ESC
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
ESC