CALIBRATION PROCEDURE
23
CALIBRATION PROCE
DURE
Chapter
4
4-2.
Procedure of equivalent input calibration and
calibration protect (torque)
The equivalent input calibration and the calibration protect uses the following procedure :
Turn off the calibration protect that inhibits
calibration.
Set the voltage input according to the torque meter
(UTM
Ⅲ
, UTF, UTM
Ⅱ
) used.
Set the unit according to the torque meter
(UTM
Ⅲ
, UTF, UTM
Ⅱ
) used.
Set the decimal place.
Set the desired minimum value of digital
increments.
(This step may be omitted if there is no change.)
Set the zero point of the sensor in no-load
condition (with the sensor unloaded).
Equivalent input calibration
Register the rated output value of the sensor,
and the capacity.
Actual load calibration
Set the span (gain) point with a load applied to
the sensor.
The calibrated value can be offset in advance.
When not using, set "0."
(This step may be omitted if there is no change.)
Set the load limit that allows digital zero.
(This step may be omitted if there is no change.)
Turn on the calibration protect to prevent
misoperation.
*
*
*
*
*
: It is indispensable.
①
Calibration protect
⑤
Min. scale division (torque)
④
Decimal place (torque)
⑧
Digital offset
⑨
Digital zero limit
⑩
Calibration protect
⑦
a. Equivalent input
calibration
③
Unit (torque)
⑥
Zero calibration
⑦
b. Actual load
calibration
or
*
②
Voltage input select