72
8
STANDARD INTERFACE
72
STANDARD INTERF
ACE
Chapter
8
■
Read setting values from UTM/UTF
Read the following setting values from UTM/UTF.
* Only applicable when "0: RS-485 command mode" "1: RS-485 display mode" is selected in
communication mode.
《
Setting value
》
0: Cancel 1: Filter setting 2: Rotation speed filter 3: Minimum rotation speed display
1.
Select setting mode 6.
Press
→
five times.
2.
Set "UTM/UTF setting value read."
Press
twice.
⇒
" UTM/UTF setting value read"
is displayed.
Press
, and present set value blinks.
Input the setting value with
.
Press
to validate it. (Initial value: 0)
3.
Press
twice to go back to the Indicated value display.
■
Write filter setting to UTM/UTF
Write the following filter setting to UTM/UTF.
In addition, after reading filter setting from UTM/UTF, the setting value from UTM/UTF will be
shown at this item no.
* Only applicable when "0: RS-485 command mode" "1: RS-485 display mode" is selected in
communication mode.
《
Setting value
》
0: 1 1: 3 2: 10 3: 30 4: 100 5: 300 6: 1k 7: PASS [Hz]
1.
Select setting mode 6.
Press
→
five times.
2.
Set "UTM/UTF filter setting."
Press
three times.
⇒
"UTM/UTF filter setting" is
displayed.
Press
, and present set value blinks.
Input the setting value with
.
Press
to validate it. (Initial value: 0)
3.
Press
twice to go back to the Indicated value display.
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
ESC
E.g. To read filter setting, enter 1 & validate. After reading setting from UTM/UTF,
Mode F6.2 of TM380 will return to 0. (Initial value)
If restart power, Mode F6.3 - F6.5 will display 0. (Not value from UTM/UTF)
Please note that setting read from UTM/UTF will not be kept.
Point
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
ESC