GIMA ABPM PULSE RATE MONITOR 

WITH BLUETOOTH

M-35112-EN-Rev

.2.07.16

0123

PROFESSIONAL MEDICAL PRODUCTS

Gima S.p.A. - Via Marconi, 1 - 20060 Gessate (MI) Italy

Italia: tel. 199 400 401 - fax 199 400 403

Export: tel. +39 02 953854209/221/225 - fax +39 02 95380056

[email protected] - [email protected]

www.gimaitaly.com

User Manual

ATTENTION: 

The operators must carefully read and completely understand 

the present manual before using the product.

CONTEC MEDICAL SYSTEMS CO., LTD

No.112 Qinhuang West Street, Economic & Technical

Development Zone, Qinhuangdao, Hebei Province,

PEOPLE’S REPUBLIC OF CHINA

Shanghai International Holding Corp. GmbH (Europe)

Eiffestrasse 80, 20537 Hamburg, Germany

Summary of Contents for ABPM PULSE RATE MONITOR WITH BLUETOOTH

Page 1: ...02 95380056 gima gimaitaly com export gimaitaly com www gimaitaly com User Manual ATTENTION The operators must carefully read and completely understand the present manual before using the product CON...

Page 2: ...2...

Page 3: ...e 13 3 4 System Menu 14 3 5 ABPM Working Interface 23 Chapter 4 NIBP MEASURING 25 4 1 Introduction 25 4 2 NIBP Measuring 25 4 3 Operation Hints 27 4 4 NIBP Error Message and Explanations 29 4 5 Mainte...

Page 4: ...Edit Trend Graph 40 6 11 Edit of Diagnose Information 42 6 12 Set BP Thresholds 43 6 13 Histogram 44 6 14 Pie Chart 44 6 15 Set Print Report 45 6 16 Print Report Preview 46 6 17 Help 48 6 18 Exit 48 C...

Page 5: ...s of strong electromagnetic interference such as radio transmitters mobile telephones etc Keep them far away high level electromagnetic radiation emitted from such devices may greatly affect the instr...

Page 6: ...ce Note The software was developed per IEC60601 1 4 The possibility of hazards arising from errors in the software program is minimized Note The monitor shall comply with the standard EN1060 series Pa...

Page 7: ...he current time date power alarm and so on User can power on off the monitor start manual measure set system parameters and so on with five keys in the front panel Please refer to Button Functions par...

Page 8: ...m Menu and left key in the trend chart The text in the right bottom of the screen indicate the function of this key Such as the button is the data review key of current user in the boot strap interfac...

Page 9: ...nect the data line to upload data 1 4 Accessories 1 A cuff for adult 2 A USB data line 3 A disk PC software 4 BP extending tube 5 A pouch Warning Please use the special accessories supplied by the man...

Page 10: ...supplied with two AA alkaline batteries or high capacity Before using the instrument you shall put the battery in the battery box in the back of the monitor Note When you don t use the equipment you s...

Page 11: ...ff LCD and enter into standby mode if there is no any operation in the standby mode the device will automatically turn off the RUN indicator flashes once every 3 seconds to prompt the device in workin...

Page 12: ...ecord will be overwritten Over flow message is shown in the boot strap interface The interface is shown as the follow 3 2 Measuring Interface Measuring interface displays real time cuff pressure and t...

Page 13: ...t press key to exit current interface and back to the boot strap interface 3 3 Measure Result Interface The measure result includes SYS systolic blood pressure mmHg kPa DIA diastolic blood pressure mm...

Page 14: ...ystem Menu In the boot strap interface according to the text in the middle bottom of the screen press button then enter the system menu and execute different option operations by using keys ALARM SOUN...

Page 15: ...ODE item has three options adult pediatric neonatal BACKLIGHT TIME s item has four choices 15 30 60 120 Note Backlight Time in the System Setup is used by the common user backlight time in ambulatory...

Page 16: ...p Select Confirm after time setup complete time change is successful and exit system time setup and return the previous menu Select Exit to cancel the setting and return to the previous menu 3 4 3 Dat...

Page 17: ...e measure result is higher than the high limit or lower than the low limit and meanwhile the ALARM SOUND is on SYS ALARM or DIA ALARM accordingly on the alarm will occur The adjustable ranges of the h...

Page 18: ...rement interval 30 minutes Awake measurement interval 15 minutes Awake time 7 00 SYS Alarm switch close DIA Alarm switch close 3 4 5 ABPM Menu Select ABPM menu in System Menu the following menu will p...

Page 19: ...ess button clear the ambulatory blood pressure measure data and enter ambulatory Blood Pressure Mode Press button save ambulatory Blood Pressure measure data and enter ambulatory Blood Pressure Mode P...

Page 20: ...nd display content include the current user total of the current user record data serial number of the record stored time of the record high pressure low pressure mean pressure Pulse Rate 3 4 6 2 ABPM...

Page 21: ...up Trend interface can trace 100 record trends if measuring data are more than 100 items press buttons can glide trend curve for left and right the scale of the vertical axis and the starting point e...

Page 22: ...interface the following interface will pop up Display content is similar to ambulatory blood pressure large print data review 3 4 7 2 Ordinary User Data Table Press button to pop up the ordinary user...

Page 23: ...table in the ordinary user data table the interface is as shown Display content is similar to ambulatory blood pressure large print data review 3 5 ABPM Working Interface In ABPM working environment...

Page 24: ...ary user working environment and display the boot strap interface Press button to exit the interface and return the ABPM working interface Long press button turn off device hint interface will pop up...

Page 25: ...to measuring mode menu setting and use children and newborn special cuff It may be dangerous for using error patient mode because the upper adult blood pressure level does not apply the children and n...

Page 26: ...isposable cuff for neonate Make sure that the cuff edge falls within the range of mark If it does not use a larger or smaller cuff that fits better 3 Connect the cuff to the air hose The limb chosen f...

Page 27: ...time interval Warning Prolonged non invasive blood pressure measurements in Auto mode may be associated with purport ischemia and neuropathy in the limb wearing the cuff When monitoring a patient exam...

Page 28: ...f the arterial pressure pulses In addition the measurement time will be prolonged Cardiac Arrhythmia s Measurements will be unreliable and may not be possible if the patient s cardiac arrhythmia has c...

Page 29: ...to dry thoroughly after washing then reinsert the rubber bag Explanation Cause Self test failure A D sampling error Loose cuff Cuff is not connected correctly Air leakage Air leakage in the valve or...

Page 30: ...s are intended for one patient use only Do not use the same cuff on any other patient Do not sterilize or use autoclave on disposable cuffs Disposable cuffs can be cleaned using soap solution to preve...

Page 31: ...ual booklet SYS Systolic pressure MAP MAP pressure DIA Diastolic pressure PR Pulse rate bpm ADU Adult PED Pediatric NEO Neonatal ABPM Ambulatory Blood Pressure Monitor INFO Information Open the alarm...

Page 32: ...e instructions in the screen Chapter 6 INTRODUCTION TO THE SOFTWARE 6 1 Main interface Double click software by left key of the mouse to enter the main operation interface as follows 1 Menu bar main o...

Page 33: ...ture above Please read the Matters need before use 6 3 Set File Path Downloaded case data will be recorded in the case path If you want to change the case path you can click menu item Set file path Af...

Page 34: ...finish the collection according to the setting Parameter explanation is as follows Patient Name the patient s name Patient ID the patient s ID number It is used for marking patient and it is exclusive...

Page 35: ...ut WIRELESS connection way When uploading the choice dialog box will appear to prompt patient to select connection way If you pitch on do not show this next time this dialog box will not appear when u...

Page 36: ...nload then select connection way After right connecting begin to download data The schedule display is as follows After finishing downloading prompt dialog box will appear and the case data will be sa...

Page 37: ...are sure that some cases don t be needed you can delete then Click menu item Delete data and deleting case interface will appear which is similar to the Select file to open interface Display as follo...

Page 38: ...interface will appear as follows The copying case method is similar to the deleting file method You can select multifile at the same time After selecting the file click Okay and Select destination di...

Page 39: ...back of is low pressure value The unit is mmHg PR BPM pulse rate value The unit is BPM MAP mmHg Mean pressure The unit is BPM PP mmHg Pressure difference between high and low blood pressure The unit i...

Page 40: ...graph shown the blue line is PR change curve and the green area is the data area below threshold and the red area is the data area above threshold The display effect of color filler type trend graph i...

Page 41: ...s move the mouse to the time axis area press left button of the mouse down dragging the mouse rightwards to stretch the BP trend graph Of course you can also drag the mouse leftwards to shrink the tre...

Page 42: ...following items Patient information Current Medications Diagnose Information and Physician information Select Patient information and display as follows Figure 45 Stretch Trend Graph Y Axis Figure 6 1...

Page 43: ...you select the Factory Defaults button Day BP load value and Night BP load value is important guide line which is used for knowing ponderance about heart blood vessel effect BP load heightening is im...

Page 44: ...only daytime statistics Night Show only nighttime statistics 6 14 Pie Chart Click shortcut key and the following interface will appear The interface is divided into four areas The most left area is m...

Page 45: ...information and so on the software will create a series of analysis reports You can select all pages or some of them for printing From the view select Edit and select Configure Report then the followi...

Page 46: ...ams graph of PR MAP PP data Measure Data Page display all BP PP PR comment and so on 6 16 Print Report Preview When you are sure that you want to print the report click shortcut key or from menu selec...

Page 47: ...In measure Data Page if the time is night the background is gray and if the Sys or Dia is beyond threshold the background is red Figure 52 Demographics Page Preview Figure 53 Measure Data Page Previe...

Page 48: ...lp Click shortcut key to pop up help file It is the brief explanation for program function 6 18 Exit Click shortcut key to save configuration information and exit application Figure 54 Measure Data Pa...

Page 49: ...rtain air tubing is not crimped 86 Measurement cancelled Push START STOP button to start reading 87 Cuff leak Check air hose and cuff 88 Safety pause Retry reading push START STOP button If problem pe...

Page 50: ...eonatal mode 145 5mmHg Resolution Pressure 1mmHg Pulse Rate 2 bpm or 2 select larger Measurement Accuracy Cuff Pressure Accuracy 3mmHg Error The BP Value of the device is equivalence with the measurem...

Page 51: ...o a low voltage power supply network which supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity The device is intended for use in the electrom...

Page 52: ...th in the location in which the device is used exceeds the applicable RF compliance level above the device should be observed to verify normal operation If abnormal performance is observed additional...

Page 53: ...MHz to 2 5 GHz 0 01 0 12 0 23 0 1 0 37 0 74 1 1 17 2 33 10 3 69 7 38 100 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m...

Page 54: ...54...

Page 55: ...55...

Page 56: ...of the warranty GIMA will repair and or replace free of charge all the defected parts due to production reasons Labor costs and personnel traveling expenses and packaging not included All components s...