![background image](http://html1.mh-extra.com/html/gima/abpm-pulse-rate-monitor-with-bluetooth/abpm-pulse-rate-monitor-with-bluetooth_user-manual_2227784049.webp)
49
Chapter 7
TROUBLESHOOTING GUIDE
Cod Description on in Report Editor Solution
1 No signal
Check position of cuff, tighten cuff
2 Overreach movement
Remain still during BP reading
4 Measurement timeout
Check air hose connections and make
certain cuff is tight
85 Airway obstructed
Check air hose connections and make
certain air tubing is not crimped
86 Measurement cancelled
Push START/STOP button to start reading
87 Cuff leak
Check air hose and cuff
88 Safety pause
Retry reading, push START/STOP button.
If problem persists return monitor
for servicing
89 Cuff overpressure
Check for blocked or kinked air hose
90 Battery low
Replace batteries. If problem persists return
for servicing
105 Self-check failed
Return for servicing
110 Out of range
Retry again, if problem persists return
for servicing
115 System error
Return for servicing