3.4.4 Alarm Setup
Select “ALARM SOUND” item in [SYSTEM MENU] to switch on or off the
sound of alarm. [refer to Figure 9 system menu]. Select “ALARM SETUP”
item in [SYSTEM MENU], the following menu will pop up:
“SYS ALARM” and “DIA ALARM” can control closing or unsealing of the SYS
and DIA alarm separately.
The alarm is on or off according to the high and low limits which have been
set up. When the measure result is higher than the high limit or lower than the
low limit, and meanwhile, the “ALARM SOUND” is on, “SYS ALARM” or “DIA
ALARM” accordingly on, the alarm will occur.
The adjustable ranges of the high and low limits of the adult mode alarm
are as follows:
SYS ALM: 40~270 mmHg
DIA ALM: 10~215 mmHg
The adjustable ranges of the high and low limits of the pediatric mode alarm
are as follows:
SYS ALM: 40~200 mmHg
DIA ALM: 10~150 mmHg
The adjustable ranges of the high and low limits of the neonatal mode alarm
are as follows:
SYS ALM: 40~135 mmHg
DIA ALM: 10~100 mmHg
Figure 14
Alarm setup