Disconnection and Storage
7-2
ULVAC CRYOGENICS INCORPORATED
Loosen the union nut using two spanners and disconnect the flexible
hose by hand.
Figure 7-1 Disconnecting Flexible Hose
7.2 Storage
Do not expose the cryopump to atmosphere for a long time. It may adsorb moisture in the
atmosphere and result in performance degradation.
Loosen the nut
with a spanner.
Hold with a
spanner.
CAUTION
◆
If the cryopump is stored in the state attached to your vacuum system, the cryopump
should be kept in a vacuum (of 100Pa or less) or under the atmospheric pressure
replaced with dry nitrogen.
◆
If the cryopump is removed from your vacuum system,
1. After the cryopump has reached room temperature completely, disconnect the
flexible hoses.
Summary of Contents for CRYO-U Series
Page 2: ......
Page 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 24: ...This page intentionally left blank S 12 ULVAC CRYOGENICS INCORPORATED...
Page 26: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 36: ...Cryopump Description 1 10 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 50: ...Installation 4 10 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 64: ...Regeneration 6 8 ULVAC CRYOGENICS INCORPORATED This page intentioanlly left blank...
Page 68: ...Disconnection and Storage 7 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 82: ...Maintenance 8 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 90: ...Troubleshooting A 8 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 92: ...Flexible Hose B 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 104: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...