Troubleshooting
Most of breakdowns of CRYO-U
®
cryopump come to light as a rise in temperature
and/or pressure through the following two routes.
(1) Refrigerator broken down
→
Cryogenic performance lowers
→
Temperature rises
→
Pressure rises
(In case of the breakdown of the refrigerator, the rise in temperature or pressure
will rapidly occur in minutes or hours)
(2) 80K shield and 15K cryopanel contaminated
→
Thermal load increases
→
Temperature rises
→
Pressure rises
(In this case, the rise in temperature or pressure will usually slowly occur in
months or years according to the use environment.)
Regarding various potential failures, the item
Ⅰ
to
Ⅶ
in Table A-1 explains its
cause and measure respectively. When the trouble cannot be solved, please contact
our Service Engineering Division or the nearest customer support center.
Summary of Contents for CRYO-U Series
Page 2: ......
Page 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 24: ...This page intentionally left blank S 12 ULVAC CRYOGENICS INCORPORATED...
Page 26: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 36: ...Cryopump Description 1 10 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 50: ...Installation 4 10 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 64: ...Regeneration 6 8 ULVAC CRYOGENICS INCORPORATED This page intentioanlly left blank...
Page 68: ...Disconnection and Storage 7 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 82: ...Maintenance 8 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 90: ...Troubleshooting A 8 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 92: ...Flexible Hose B 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 104: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...