CRYO-U
®
Cryopump Instruction Manual
ULVAC CRYOGENICS INCORPORATED
1-3
1.3 Environmental Requirements
Condition
Requirement
Temperature when in
operation(*1)
10 - 35 °C
Temperature when not in
operation
-10 - 55 °C
Humidity (*1)
5
~
85% (non-condensing)
Altitude
< 1000m
Atmosphere
Free from metallic powder, dust, combustible gas, or
corrosive gas
(*) Depending on the ambient temperature and humidity, dew condensation may form on
an operating cryopump as the surface of the pump case becomes cooler than the
dew-point (Refer to “Appendix F Dew Point Temperature”). Use an air conditioner to
control temperature and humidity to prevent condensation.
◆
Thermometer
・
MBS-C CRYO METER
・
Hydrogen Vapor Pressure Gauge
Charged Pressure
:
0.35
±
0.02MPaG at room temp.
・
K thermocouple
Electromotive Force
:
Approximate 0mV at 20ºC
◆
Accessory Flange(C-type)
φ
4
4
.
5
φ
34.9
φ
2
2
2
.
1
7
1.25
7
O ring
AS568-020
4-φ4.5
hole
Summary of Contents for CRYO-U Series
Page 2: ......
Page 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 24: ...This page intentionally left blank S 12 ULVAC CRYOGENICS INCORPORATED...
Page 26: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 36: ...Cryopump Description 1 10 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 50: ...Installation 4 10 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 64: ...Regeneration 6 8 ULVAC CRYOGENICS INCORPORATED This page intentioanlly left blank...
Page 68: ...Disconnection and Storage 7 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 82: ...Maintenance 8 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 90: ...Troubleshooting A 8 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 92: ...Flexible Hose B 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 104: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...