Model Two & Model Two Lite - Operation Manual
67
If you wish the Auto Test program to perform its test sequence in programmed time intervals,
press
F2
to setup the 24 HR Measurement routine.
The Auto Test screen for STEP 5 appears.
A default time interval of 6 hours between test times is displayed. To change the time interval,
press
FCN
, enter a new number from 1 to 23 (hours), and press
F1
to enter.
Press
F2
(NEXT) to continue. The Auto Test screen for STEP 6 appears.
A default number of 5 test times is displayed. To change the number of test times, press
FCN
,
enter a new number from 1 to 10 (times), and press
F1
to enter.
Note:
Since the first Auto Test time is always performed immediately
upon starting the program, the 5 default test times provides a 24 Hour
span from first test to last test, with a 6-hour default time interval between
test sequences.