C-Link Protocol Commands
Alarms
B-18
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
Receive:
set alarm sample rh max b 40.0 ok
alarm status det a/b
This command reports the board status of the detector as FAIL or OK.
The following example reports that the detector board status is OK.
Send:
alarm status det a
Receive:
alarm status det a ok
Send:
alarm status det b
Receive:
alarm status det b ok
alarm tape counter max
This command reports the current filter tape count alarm maximum
value setting. The following example reports that the filter tape count
alarm maximum value is 480.
Send:
alarm tape counter max
Receive:
alarm tape counter max 480
set alarm tape counter max
value
This command sets the filter tape count alarm maximum
value
, where
value
is an integer number representing filter tape alarm limits. The
following example sets the filter tape count alarm maximum value to
490.
Send:
set alarm tape counter max 490
Receive:
set alarm tape counter max 490 ok
alarm vac pres min a/b
alarm vac pres max a/b
These commands report the current vacuum pressure alarm minimum
and maximum value settings. The following example reports that the
vacuum pressure alarm minimum value is -20.0 mmHg.
Send:
alarm vac pres min a
Receive:
alarm vac pres min a -20.0 mmHg
Send:
alarm vac pres min b
Receive:
alarm vac pres min b -20.0 mmHg
set alarm vac pres min a/b
value
set alarm vac pres max a/b
value
These commands set the vacuum pressure alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing vacuum pressure alarm limits in millimeters of mercury.
The following example sets the vacuum pressure alarm maximum value
to 250.0 mmHg.
Send:
set alarm vac pres max a 250.0
Receive:
set alarm vac pres max a 250.0 ok
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......