Operation
Service Menu
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
3-73
The Calibrate Barometer Pressure Span screen allows the user to view
and pressure sensor calibration span point.
Note
The operator should use an independent barometer to measure
the ambient pressure and enter the value on this screen before
calibrating.
s
l
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Baro Pres Calibration >
Span
.
CAL BARO PRESSURE SPAN:
PRESSURE: 760 mmHg
SET TO: 74
7
mmHg?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Restore Default Calibration screen allows the user to reset the
default values.
l
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Baro Pres Calibration >
Set Defaults
.
RESTORE DEFAULT CAL:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
The Vacuum/Flow or Vaccum/Flow B Calibration submenu are used
to calibrate the vacuum/flow zero, vacuum/flow span, or reset default
values.
Note
Vac/Flow B settings please refer to Vac/Flow as following
description.
s
l
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Vac/Flow or Vac/Flow B
.
VAC PRES CALIBRATION:
>VAC/FLOW OFFSET
VAC PRES SPAN 1.0000
FLOW PRES SPAN 1.0000
SET DEFAULTS
RANGE AVG DIAGS ALARM
Calibrate Barometer Pressure
Span
Restore Default Calibration
Vacuum/Flow or Vacuum/ Flow B
Calibration
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......