Operation
Alarms Menu
3-66
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
BAROMETRIC PRESSURE:
ACT MIN: 400.0 mmHg
SET MIN: 400.1 mmHg?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Vacuum screen displays the current vacuum reading and sets the
minimum and maximum alarm limits. Acceptable alarm limits range
from -5 mmHg to 250 mmHg. If the vacuum reading goes beyond the
minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu
Note
Vacuum B settings please refer to Vaccum as following
description.
s
l
In the Main Menu, choose Alarms > Pressure/Vacuum Alarms >
Vacuum or Vacuum B
.
VACUUM:
ACTUAL 60.8 mmHg
>MIN -5.0 mmHg
MAX 250.0 mmHg
RANGE AVG DIAGS ALARM
The Minimum Vacuum alarm limit screen is used to change the
minimum vacuum alarm limit. The minimum and maximum vacuum
screens function the same way.
Note
The maximum vacuum value should not be changes, as this is also
used as a protective feature for the instrument.
s
l
In the Main Menu, choose Alarms > Pressure/Vacuum Alarms >
Vacuum or Vacuum B >
Min
or
Max
.
VACUUM:
ACT MIN: -5.0 mmHg
SET MIN: -4.9 mmHg?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Vacuum
Min and Max Vacuum Limits
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......