Servicing
Service Locations
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
7-35
2. Disconnect the ribbon cable and the two-wire connector from the
front panel board.
3. Remove the four screws at the corners of the LCD module.
4. Slide the LCD module out towards the center of the instrument.
5. Replace the LCD module by following the previous steps in reverse.
Note
The optimal contrast will change from one LCD screen to
another. After replacing the LCD screen, the contrast may need to be
reset. If the content on the screen is visible, select Instrument Controls
>
Screen Contrast
and adjust the screen contrast. If the content on
the screen is not visible, use the “set contrast 10” C-Link command to
set screen constrast to mid range, then optimize the contrast. See the
” appendix for more information on this
command.
s
For additional assistance, service is available from exclusive distributors
worldwide. Contact one of the phone numbers below for product
support and technical information or visit us on the web at www.
thermofisher.com.
1-866-282-0430 Toll Free
1-508-520-0430 International
Service Locations
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......