The Analog Output Calibrate Zero screen allows the user to calibrate
the zero state of the selected analog output. The operator must connect
a volt meter to the output and adjust the output until it reads 0.0 V
on the meter for a voltage channel, or either 0 or 4 mA for a current
channel (depending on the selected range). See the “set output to:” field
on the display.
l
In the Main Menu, choose Service > Analog Out Cal > select
Channel >
Calibrate Zero
.
ANALOG OUTPUT CAL: ZERO
CONNECT METER TO OUTPUT!
SELECTED OUTPUT: V1
SET TO: 100
SAVE VALUE
INC/DEC
SET OUTPUT TO: 0.0 V
RANGE AVG DIAGS ALARM
The Analog Output Calibrate Full-Scale screen allows the user to
calibrate the full-scale state of the selected analog output. The operator
must connect a volt meter to the output and adjust the output until the
meter reads the value shown in the “set output to:” field, in either V or
mA, depending on the selected output channel.
l
In the Main Menu, choose Service > Analog Out Cal > select
Channel >
Calibrate Full Scale
.
ANALOG OUTPUT CAL: SPAN
CONNECT METER TO OUTPUT!
SELECTED OUTPUT: V1
SET TO: 3397
SAVE VALUE
INC/DEC
SET OUTPUT TO: 10 V
RANGE AVG DIAGS ALARM
The Analog Input Calibration menu is used to calibrate the 8 analog
input channels at both zero and full-scale. The analog input calibration
menu is visible only when the I/O expansion board is installed and
when the instrument is in service mode. For more information on the
Note
This adjustment should only be performed by an instrument
service technician.
s
l
In the Main Menu, choose Service >
Analog Input Calibration
.
Analog Output Calibrate Zero
Analog Output Calibrate Full-
Scale
Analog Input Calibration
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......