C-Link Protocol Commands
Measurement Configuration
B-42
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
set flow target a/b
value
This command sets the flow to
value
, where
value
is a floating-point
number representation of flow between 1 to 20 LPM. The following
example sets the flow to 16.67 LPM.
Send:
set flow target a 16.67
Receive:
set flow target a 16.67 ok
Send:
set flow target b 16.67
Receive:
set flow target b 16.67 ok
ht control a/b
This command reports the current status of the heater (off, RH
threshold, or temp threshold). The following example reports that the
heater is RH threshold.
Send:
ht control a
Receive:
ht control a RH
Send:
ht control b
Receive:
ht control b RH
set ht control
selection
a/b
selection
= | off | rh | temp |
This command sets the heater to off, RH threshold, or temp threshold.
The following example sets the heater to temp threshold.
Send:
set ht control temp a
Receive:
set ht control temp a ok
Send:
set ht control temp b
Receive:
set ht control temp b ok
mass limit
This command reports the current filter mass limit. The following
example reports that the mass limit is 1,500 mg.
Send:
mass limit
Receive:
mass limit 1500
set mass limit
value
This command sets the filter mass limit to
value
, where
value
is a
floating-point number representation of mass limit between 0.5 to 5.0
mg or 500 to 5,000 µg. The following example sets the mass limit to 1.0
mg.
Send:
set mass limit 1.0
Receive:
set mass limit 1.0 ok
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......