C-Link Protocol Commands
Datalogging
B-28
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
Send:
malloc lrec
Receive:
malloc lrec 10
%
set malloc lrec
value
set malloc srec
value
value
= 0 to 100
These commands set the percent of memory space allocated for lrecs
and sreco to value, where value is a floating-point number representing
percent. The following example sets the memory allocation for lrecs to
10%.
Note
Issuing these commands will clear all the logged data in
memory. All the existing records should be retrieved using appropriate
commands, if required.
s
Send:
set malloc lrec 10
Receive:
set malloc lrec 10 ok
no of lrec
no of srec
These commands report the number of lrec and srecs stored memory.
The following example shows that 50 lrecs have been stored in the
memory.
Send:
no of lrec
Receive:
no of lrec 50 recs
sp field
number
This command reports the variable
number
and name stored at the
index in the scratch pad list.
The scratch pad is a temporary memory area which is used to set up lists
of selections for lrec, srec, or streaming data items. The user can copy
any of these lists to the scratch pad, modify individual elements in the
list, then save the scratch pad back to the original list.
The following example shows that field 1 in the scratch pad is set to
index number 3, which is for the variable PM concentration pressure.
Send:
sp field 1
Receive:
sp field 1 3 PM
set sp field
number value
number
= 1-32 for lrec and srec lists, 1-8 for streaming data lists
This command sets the scratch pad field
number
(item number
in scratch pad list) to
value
, where
value
is the index number of a
variable in the analog output variable list. Available variables and their
corresponding index numbers may be obtained using the command “list
var aout”. The “set sp field” command is used to create a list of variables
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......