C-Link Protocol Commands
Communications Configuration
B-52
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
Note
If DHCP is changed from on to off and then the IP address, the
netmask address, or the gateway address is changed, you must cycle
power to the instrument before the change takes effect. Until you cycle
the power, the address assigned by the DHCP server will still be used
and reported as the current address.
s
Send:
set dhcp on
Receive:
set dhcp on ok
format
This command reports the current reply termination format. The
following example shows that the reply format is 00, which means reply
with no checksum, according to
.
Send: format
Receive:
format 00
set format
format
This command sets the reply termination
format
, where
format
is
set according to
. The following example sets the reply
termination format to checksum.
Send:
set format 01
Receive:
set format 01 ok
Table B–8.
Reply Termination Formats
Format
Reply Termination
00
<CR>
01
<NL> sum xxxx <CR>
where xxxx = 4 hexadecimal digits that represent the sum of all the
characters (bytes) in the message
host name
This command reports the host name string. The following example
reports the host name is set to iSeries.
Send:
host name
Receive:
host name iSeries
set host name
string
This command sets the host name
string
, where
string
is 1-13
alphanumeric characters. The following example sets the host name to
analyzer01.
Send:
set host name analyzer01
Receive:
set host name analyzer01 ok
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......