Troubleshooting
Troubleshooting Guides
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
6-5
Alarm Message
Possible Cause
Action
RH/Temperature Alarms –
Ambient RH
Disconnect cable
(external) on rear
panel
Verify cable connection.
Disconnect cable
(internal)
Verify cable connection.
RH/Temperature Alarms –
Sample RH
Disconnected heater Verify cable connection.
RH/Temperature Alarms –
Ambient Temp
Disconnect cable
(external) on rear
panel
Verify cable connection.
Disconnect cable
(internal)
Verify cable connection.
RH/Temperature Alarms –
Flow Temp
Disconnected cable
Verify connection to measurement
interface board.
Sensor defective
Replace sensor.
RH/Temperature Alarms –
Board Temp
Disconnected cable
Verify connection to measurement
interface board.
Sensor defective
Replace sensor.
Pressure/Vacuum Alarms
– Barometer Pressure
Disconnected cable
Verify cable connections.
Replace board.
Pressure/Vacuum Alarms
– Vacuum
Disconnected
plumbing
Verify plumbing connections.
Disconnected cable
Verify cable connection.
Sensor defective
Replace pressure board.
Pressure/Vacuum Alarms
– Flow
Disconnected
plumbing
Verify plumbing connections
Disconnected cable
Verify cable connection.
Sensor defective
Replace pressure board.
D
P valve open
Verify valve is in close position.
Flow Alarms – Flow
Low flow
Check that pump is connected.
Verify that both
D
P valves are in the
closed position.
No flow
Verify proportional valve plumbing.
Verify connection to measurement
interface board.
Conc Alarms – Inst PM
Conc negative
Check heater and Beta counts.
Conc alarm setting
Verify conc alarm settings.
Conc Alarms – Avg PM
Conc negative
Check heater and Beta counts.
Conc alarm setting
Verify conc alarm settings.
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......