Operation
Instrument Controls Menu
3-18
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
l
In the Main Menu, choose Instrument Controls >
Filter Tape
Control
.
FILTER TAPE CONTROL:
>CHANNEL A
CHANNEL B
RANGE AVG DIAGS ALARM
CHANNEL:
>MANUAL
MASS LIMIT 1,500
µ
g
NEXT TIME 05MAY08 00:00
PERIOD 8 HRS
COUNTER 0
RANGE AVG DIAGS ALARM
The Manual screen allows the user to move or stop the filter tape and
open or close the measurement bench.
Note
Channel B settings please refer to Channel A as following
description.
s
l
In the Main Menu, choose Instrument Controls > Filter Tape
Control > Channel A or Channel B>
Manual
.
MANUAL:
>BENCH(S) CLOSED
TAPE STOP
TOGGLE VALUE
RANGE AVG DIAGS ALARM
The Mass Limit screen allows the user to set the mass limit. The default
is 1,500 µg (or 1.5 mg) and is specified as part of any U.S. EPA and
TUV approvals. However, the limit can be increased as much as 5,000
µg for non-regulatory monitoring.
Note
Channel B settings please refer to Channel A as following
description.
s
l
In the Main Menu, choose Instrument Controls > Filter Tape
Control > Channel A orChannel B>
Mass Limit
.
Manual
Mass Limit
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......