Installation
Acceptance Testing and Startup Procedures
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
2-5
the filter tape. The flow pressure is the pressure differential across the
orifice.
Table 2–1.
Pressure Units Conversion
Pressure Conversions
mmHg = atm x 0.0013158
mmHg = inHg x 0.0393701
mmHg = hPa x 1.3332237
Assuming adequate time has passed for thermal equilibration
(approximately 1 hour), and the vacuum pump has been drawing room
air into the instrument, compare the ambient temperature, sample
temperature, and flow temperature to your NIST traceable thermometer
placed at the inlet tube (see
).
Note
Please be sure that the pump exhaust or any other heat source
is not influencing the sensor reading of the temperature/RH cable
assembly.
s
Figure 2–1.
Model 5028
i
Sampling Room Air Under Stable Conditions
Record the temperature sensor readings from the Model 5028
i
and compare to your NIST-traceable thermometer. Each of these
measurements should be within ±2 °C tolerance of your NIST-traceable
thermometer. Within this tolerance, the temperature sensors have
passed the acceptance test.
If the sensors are slightly out of tolerance (±3 °C), the acceptance test
should be classified as marginal. If the sensor performance is outside this
range, please perform a temperature sensor calibration.
One-Point Temperature
Verification
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......