Operation
Alarms Menu
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
3-69
The Average PM screen displays the current average PM concentration
and sets the minimum and maximum alarm limits. Acceptable alarm
limits range from 0 to 10000 µg/m
3
or 0 to 10 mg/m
3
. If the average
PM concentration goes beyond either the minimum or maximum limit,
an alarm is activated and the alarm (bell) icon appears in the status bar
on the Run screen and in the Main Menu.
Note
Avg PM
b
settings please refer to Avg PM as following description.
s
l
In the Main Menu, choose Alarms > Conc Alarms >
Avg PM/Avg
PM
b
.
AVERAGE PM CONC:
ACTUAL 0.0
>MIN 0.0
MAX 10000
RANGE AVG DIAGS ALARM
The Minimum Average PM/PM
b
Concentration alarm limit screen is
used to change the minimum average PM/PM
b
concentration alarm
limit. The minimum and maximum average PM/PM
b
concentration
screens function the same way.
l
In the Main Menu, choose Alarms > Conc Alarms > Avg PM or Avg
PM
b
>
Min
or
Max
.
AVERAGE PM CONC:
ACT MIN: 0.0
SET MIN:
xxxxx
10
0
?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Instant PM screen displays the current instant PM/PM
b
concentration and sets the minimum and maximum alarm limits.
Acceptable alarm limits range from -10 to 10000 µg/m
3
or -.01 to 10
mg/m
3
. If the instant PM/PM
b
concentration goes beyond either the
minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
Note
Instant PM
b
settings please refer to Instant PM as following
description.
s
l
In the Main Menu, choose Alarms > Conc Alarms >
Inst PM or
Inst PM
b
.
Average PM/PM
b
Min and Max Average PM/PM
b
Concentration Limits
Instant PM/Instant PM
b
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......