Installation
Data Content
2-18
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
5-minute logging with 70% memory allocated. If there are
any changes, be sure to Commit Content.
b. From the Main Menu, choose Service. Under Service:
Note
If Service Mode is not displayed, refer to “
, then return to the beginning of this step.
s
i. Check/Calibrate Ambient Temperature, Sample Temperature,
and Flow Temperature
ii. Check/Calibrate Ambient RH
iii. Check Calibrate Barometric Pressure Span
iv. Check/Calibrate Flow Calibration
v. Perform a Auto Detector Calibration (30 minutes)
vi. Perform a Mass Foil Calibration (factory calibrated)
vii. Mass Coefficients are set at the factory.
Please use iPort to access data download from the instrument.
Use the following procedure when shutting down the instrument.
1. Turn the power OFF.
2. Cover instrument ports when not in use.
DO NOT
l
Do not pressurize the monitor.
l
Do not allow liquids to enter the sample port.
l
Do not expose the instrument to excessive vibration or magnetic
interference.
l
Do not remove/adjust the beta attenuation bench of the 5028
i
.
l
Do not remove the C-14 radioactive source at any time.
DO
l
Operate the instrument in a climate-controlled environment
(4–50 °C).
l
Set DHCP to ON when changing the IP address.
Data Content
Shutdown
Important Tips
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......