Gesytec (Bayern-Hessen) Protocol
Gesytec Commands
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
D-7
The following 9 measurements reported in response to DA command
for the Model 5028
i
include:
l
PM
l
Mass
l
Volumetric Flow
l
Ambient Temperature
l
Flow Temperature
l
Ambient RH
l
Flow Pressure
l
Vacuum Pressure
l
Barometric Pressure
See
for error status for
the Model 5028
i
.
Table D–1.
Operating Status for Model 5028
i
D7
D6
D5
D4
D3
D2
D1
D0
è
Bit
8
7
6
5
4
3
2
1
à
Hex-value
80
40
20
10
08
04
02
01
MSB
LSB
Operating status:
Service Mode (On)
0
0
0
0
0
0
0
1
Maintenance (Local)
0
0
0
0
0
0
1
0
Pump (Off)
0
0
0
0
0
1
0
0
Heater (Off)
0
0
0
0
1
0
0
0
Filter Change
0
0
0
1
0
0
0
0
Not Used
0
0
1
0
0
0
0
0
Not Used
0
1
0
0
0
0
0
0
Not Used
1
0
0
0
0
0
0
0
Measurements reported in
response to DA command
Operating and Error Status
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......