Operation
Alarms Menu
3-68
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
The Flow screen displays the current flow and sets the minimum and
maximum alarm limits. Acceptable alarm limits range from 16.00
to 17.34 LPM. If the flow reading goes beyond the minimum or
maximum limit, an alarm is activated and the alarm (bell) icon appears
in the status bar on the Run screen and in the Main Menu.
Note
Flow B settings please refer to Flow as following description.
s
l
In the Main Menu, choose Alarms > Flow Alarms >
Flow or Flow B
.
FLOW:
ACTUAL 0.00 LPM
>MIN 16.17 LPM
MAX 17.17 LPM
RANGE AVG DIAGS ALARM
The Minimum Flow alarm limit screen is used to change the minimum
flow alarm limit. The minimum and maximum flow screens function
the same way.
l
In the Main Menu, choose Alarms > Flow Alarms > Flow or Flow B
>
Min
or
Max
.
FLOW:
ACT MIN: 16.00 LPM
SET MIN: 16.17 LPM?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Concentration Alarms submenu allows the user to view and set the
24-hour average PM and instant PM concentration alarm parameters.
l
In the Main Menu, choose Alarms >
Conc Alarms
.
CONC ALARMS:
>AVG PM OK
AVG PMb OK
INST PM OK
INST PMb OK
RANGE AVG DIAGS ALARM
Flow
Min and Max Flow Limits
Concentration Alarms
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......