C-Link Protocol Commands
Measurement Configuration
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
B-45
temp comp
This command reports whether temperature compensation is for actual
or standard conditions. The following example shows the temperature
compensation is standard.
Send:
temp comp
Receive:
temp comp std
set temp comp
selection
selection
= | act | std |
These commands turn the temperature compensation to either actual
or standard conditions. The following example turns the temperature
compensation to actual.
Send:
set temp comp act
Receive:
set temp comp act ok
temp std a/b
This command reports the standard temperature. The following example
reports that the standard temperature is 18 °C.
Send:
temp std a
Receive:
temp std a 18 degC
Send:
temp std b
Receive:
temp std b 18 degC
set temp std a/b
value
This command sets the standard temperature to
value
, where
value
is a
floating-point number representation of standard temperature between
0 to 25 °C. The following example sets the standard temperature to 15
°C.
Send:
set temp std a 15
Receive:
set temp std a 15 ok
Send:
set temp std b 15
Receive:
set temp std b 15 ok
temp th a/b
This command reports the heater temperature threshold for the sample.
The following example reports that the temperature threshold is 30 °C.
Send:
temp th a
Receive:
temp th a 30 degC
Send:
temp th b
Receive:
temp th b 30 degC
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......