BLOOD PRESSURE MONITOR

UPPER ARM

UA-651CEL-SM

Instruction Manual 

Summary of Contents for SMARTMETER iBloodPressure 2.0

Page 1: ...CELLULAR BLOOD PRESSURE MONITOR UPPER ARM UA 651CEL SM Instruction Manual...

Page 2: ...in temperature humidity direct sunlight shock or dust should be avoided Clean the device and cuff with a dry soft cloth or a cloth dampened with water and a neutral detergent Never use alcohol benzene...

Page 3: ...h as home networking devices mobile phones cordless phones and their base stations walkie talkies can affect this blood pressure monitor Therefore a minimum distance of 30 cm 12 should be kept from su...

Page 4: ...to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operations Thi...

Page 5: ...That may result in electrical shock Do not inflate without wrapping the cuff around the upper arm A D Medical and Smart Meter cannot accept liability for any damages incurred due to impaired operation...

Page 6: ...7 Using the Monitor E 9 Taking your Blood Pressure E 14 What is an Irregular Heartbeat E 17 Pressure Bar Indicator E 17 AHA Classification E 18 About Blood Pressure E 19 Troubleshooting E 21 Maintena...

Page 7: ...Proper fit range Artery position mark DC jack AC adapter Plug Air Connector Plug Battery Compartment Battery Cover Systolic Pressure Diastolic Pressure Battery Indicator Pressure Bar Indicator Pulse r...

Page 8: ...nks Replace all batteries with new ones when the indicator blinks Unstable blood pressure due to movement during the measurement Take another measurement Remain still during the measurement The systol...

Page 9: ...device on SYS Systolic blood pressure in mmHg DIA Diastolic blood pressure in mmHg PUL min Pulse per minute R6 LR6 AA Battery installation guide Direct current IP International protection symbol SN Se...

Page 10: ...that the battery needs to be replaced replace all batteries with new ones Do not mix old and new batteries It may shorten the battery life or cause the device to malfunction Replace the batteries two...

Page 11: ...ter part TB 233 is sold separately Use the specified AC adapter See Technical Data When disconnecting the AC adapter from the electrical outlet grasp the body of the AC adapter plug and gently pull th...

Page 12: ...ith the cellular network This monitor has no clock adjustment function Selecting the correct cuff Using the correct cuff size is important for an accurate reading If the cuff is not the proper size th...

Page 13: ...s designed to provide special protection against electrical shocks W Proper fit range for the Wide Range cuff L Proper fit range for the Large cuff M Proper fit range for the Medium cuff S Under range...

Page 14: ...directly against the skin as clothing may cause a faint pulse and result in a measurement error 2 Constriction of the upper arm caused by rolling up a shirtsleeve may prevent accurate readings 3 Confi...

Page 15: ...vel Measure two times a day in the morning and evening Notes for Accurate Measurement This monitor bases its measurements on the heartbeat If you have a very weak or irregular heartbeat the monitor ma...

Page 16: ...ts automatically and the heart mark blinks indicating that the measurement is in progress Once the pulse is detected the mark blinks with each pulse beat Note If an appropriate pressure is not obtaine...

Page 17: ...nks press and hold the START button until a number about 30 to 40 mmHg higher than your expected systolic pressure appears 4 When the desired number is reached release the START button to start measur...

Page 18: ...hythm during the measurements the IHB indicator will appear on the display with the measurement values Note We recommend contacting your physician if you see this IHB indicator frequently Pressure Bar...

Page 19: ...ion Systolic mmHg Diastolic mmHg Normal 120 and 80 Elevated 120 129 and 80 Hypertension Stage 1 High Blood Pressure 130 139 or 80 89 Hypertension Stage 2 High Blood Pressure 140 or 90 HYPERTENSIVE CRI...

Page 20: ...ure if left unattended can cause many health problems including stroke and heart attack Hypertension can be controlled by altering lifestyle avoiding stress and with medication under a doctor s superv...

Page 21: ...while at work or play and falls to its lowest levels during sleep So do not be overly concerned by the results of one measurement Take measurements at the same time every day using the procedure descr...

Page 22: ...measurement Make sure you remain very still and quiet during measurement The cuff position is not correct Sit comfortably and still Place your arm on a table with your palm facing upward and the cuff...

Page 23: ...thod Oscillometric measurement Measurement range Pressure 0 299 mmHg Systolic pressure 60 279 mmHg Diastolic pressure 40 200 mmHg Pulse 40 180 beats min Measurement accuracy Pressure 3 mmHg Pulse 5 Po...

Page 24: ...gress protection Device IP20 Applied part Cuff Type BF Useful life Device 5 years when used six times a day Cuff 2 years when used six times a day NOTE Specifications are subject to change without pri...

Page 25: ...guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equip...

Page 26: ...mpliance Conducted and radiated RF EMISSION CISPR 11 Group 1 Class B Harmonic distortion IEC 61000 3 2 Class A Voltage fluctuations and flicker IEC 61000 3 3 Compliance Table 2 IMMUNITY TEST LEVELS En...

Page 27: ...ption IEC 61000 4 11 0 UT 250 300 cycle NOTE UT is the AC mains voltage prior to application of the test level Table 4 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equ...

Page 28: ...28 1845 1970 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation 217 Hz 2 0 3 28 5240 5100 5800 WLAN 802 11 a n Pulse modulation 217 Hz 0 2 0 3 9 5500 5785 Accessories are...

Page 29: ...cts in materials and workmanship for two years from the date of purchase This warranty does not apply to the performance of an iBloodPressure 2 0 meter that has been altered misused tampered with or a...

Page 30: ...B L O O D P R E S S U R E R E C O R D R E G I S T R O D E L A P R E S I N A R T E R I A L Name Nombre Age Edad Weight Peso DATE FECHA AM SYS DIA PULSE PULSO PM SYS DIA PULSE PULSO...

Page 31: ...B L O O D P R E S S U R E R E C O R D R E G I S T R O D E L A P R E S I N A R T E R I A L Name Nombre Age Edad Weight Peso DATE FECHA AM SYS DIA PULSE PULSO PM SYS DIA PULSE PULSO...

Page 32: ...Company Limited 1 243 Asahi Kitamoto shi Saitama ken 364 8585 Japan Manufactured for Smart Meter LLC 5501 W Waters Ave STE 401 Tampa FL 33634 SmartMeterRPM com 2022 All trademarks are property of the...