ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR
cause the data to be invalid. The Purge Nozzle Assemblies need not be
removed during each maintenance; however, Teledyne Monitor Labs
recommends that the transducers be visually inspected approximately
every six months.
9.2 TROUBLESHOOTING GUIDE
One of the design features of the Ultraflow 150 system is that all of the
information necessary to troubleshoot the system to a subassembly level
can be obtained from the Enhanced Remote Panel. The Status Codes
together with the Service Data Values will define almost any system
problem.
If a Fault is indicated or you suspect a problem:
Go to the Enhanced Remote Panel and write down the status word and
Service Data values.
Decode the status word according to the help screen or tables of
section four. This alone may point to the problem.
Compare the Service Data readings to the readings from your last
maintenance check sheet.
If at this point the problem is not apparent:
Go through the maintenance check sheet procedure (Appendix B).
Review this data for the proper ranges and compare it to those of
previous maintenance.
If at this point you have not been able to determine the problem:
Call the Teledyne Monitor Labs tech support line (800) 846-6062. A
factory trained technician with years of experience will handle your
call. Please have available:
o
Instrument model and serial number
o
Data collected in the maintenance document above
9-2
Summary of Contents for ULTRAFLOW 150
Page 2: ...DOCUMENT NO 1900 0100 01 REV E ...
Page 24: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank 4 6 ...
Page 75: ...SECTION 7 0 INSTALLATION This page intentionally left blank 7 7 ...
Page 76: ......
Page 85: ...SECTION 8 0 SYSTEM CALIBRATION AND ADJUSTMENT This page intentionally left blank 8 9 ...
Page 86: ......
Page 89: ...APPENDIX A SITE SPECIFICATION DATA SHEETS ...
Page 90: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank ...
Page 95: ...APPENDIX B MAINTENANCE CHECK SHEETS ...
Page 96: ... This page intentionally left blank ...
Page 101: ...APPENDIX C SPARE PARTS ...
Page 102: ... This page intentionally left blank ...
Page 105: ......
Page 106: ...APPENDIX D DRAWINGS ...
Page 107: ... This page intentionally left blank ...
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ...APPENDIX E ENHANCED SERIAL PORT COMMUNICATION PROTOCOL ...