ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR
The WITH CAL field can be "YES", "NO" or “EXP”. Use the “Page-
up/Page-down” keys to find the parameter you want and press ENTER to
select. The action of each selection is explained below.
WITH CAL = YES:
Calibration data will be sent to that output when a
calibration request is processed. Scaling for the calibration data will be
the same as the Normal mode scaling entered under VALUE AT ZERO
SCALE and VALUE AT FULL SCALE.
WITH CAL = NO:
The output will not receive or display calibration
values on it.
WITH CAL = EXP:
Calibration data will be sent to that output when a
calibration request is processed. Scaling for the SPANHIGH and
SPANLOW calibration data will be the same as the Normal mode scaling
entered under VALUE AT ZERO SCALE and VALUE AT FULL
SCALE. However, scaling for ZERO calibration data will be ±10% of the
VALUE AT FULL SCALE.
The next field chooses the TIE/path you want the output to look at. Again
use “Page-up/Page-down” to find the path you want and press ENTER to
select and move to the next field.
The field labeled VALUE AT ZERO SCALE: is the numerical value that
will correspond to 4mA output. VALUE AT FULL SCALE: is the
numerical value at full-scale (20mA) output. These values are what you
would change to rescale the output if needed. Enter the numbers desired
through the keypad. Press
“
↵
”
to move to the next field.
The second half of this screen offers the same options for Output #2.
Make the desired choices in the same way or use the arrow keys to scroll
to the “SAVE THIS CONFIG” line. If satisfied with the selections, press
enter to save the selections you just made.
NOTE:
On this screen no changes are accepted by the Multi I/O
configuration until the last field, SAVE THIS CONFIG, is entered. This
field confirms that you want to implement the changes entered. If one of
the fields is not correct use the
↑
or
↓
key to scroll to it. Once corrected
move to “SAVE THIS CONFIG and press the enter “
↵
” key.
6.4.7.2 Analog Outputs 3,4
This screen is password protected
. The ANALOG OUTPUTS 3,4 menu
screen functions exactly the same as the ANALOG OUTPUTS 1,2 menu
screen. Only the output # references have changed. Refer to the Analog
Outputs 1,2 section for a functional description.
6-18
Summary of Contents for ULTRAFLOW 150
Page 2: ...DOCUMENT NO 1900 0100 01 REV E ...
Page 24: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank 4 6 ...
Page 75: ...SECTION 7 0 INSTALLATION This page intentionally left blank 7 7 ...
Page 76: ......
Page 85: ...SECTION 8 0 SYSTEM CALIBRATION AND ADJUSTMENT This page intentionally left blank 8 9 ...
Page 86: ......
Page 89: ...APPENDIX A SITE SPECIFICATION DATA SHEETS ...
Page 90: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank ...
Page 95: ...APPENDIX B MAINTENANCE CHECK SHEETS ...
Page 96: ... This page intentionally left blank ...
Page 101: ...APPENDIX C SPARE PARTS ...
Page 102: ... This page intentionally left blank ...
Page 105: ......
Page 106: ...APPENDIX D DRAWINGS ...
Page 107: ... This page intentionally left blank ...
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ...APPENDIX E ENHANCED SERIAL PORT COMMUNICATION PROTOCOL ...