ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR
Table 6-10
Parameters for Analog Output
NOTE: Each of the parameters for output can carry data from either TIE and either path or the
average of the two paths. The TIE and path are selected on the output configuration
screen.
Parameter Name
Functional Description
INSTANT VELOCITY
This parameter is the linearized, instantaneous, actual velocity as measured during a
single integration period. It is linearized per the curve developed from site testing
AVE VELOCITY
This parameter is an average of the linearized, instantaneous velocity values from a
number of integration periods. The number of periods is set by the “periods for
averaging” variable. The length of the averaging period will be “seconds of
integration” X “number of periods averaged”.
INSTANT RAW VELOCITY
This parameter is the raw, unlinearized, instantaneous, actual velocity as measured
during a single integration period.
AVE RAW VELOCITY
This parameter is an average of the raw unlinearized instantaneous velocity values
from a number of integration periods. The number of periods is set by the “periods
for averaging” variable. The length of the averaging period will be “seconds of
integration” X “number of periods averaged”.
INSTANT ACT VOLUME
This parameter is the linearized, instantaneous, actual volume as measured during a
single integration period. It is linearized per the curve developed from site testing .
AVE ACT VOLUME
This parameter is an average of the linearized, instantaneous actual volume values
from a number of integration periods. The number of periods is set by the “periods
for averaging” variable. The length of the averaging period will be “seconds of
integration” X “number of periods averaged”.
INSTANT STD VOLUME
This parameter is the linearized, instantaneous, standard volume as measured during a
single integration period. It is linearized per the curve developed from site testing
and corrected to standard temperature and/or pressure conditions.
AVE STD VOLUME
This parameter is an average of the linearized, instantaneous standard volume values
from a number of integration periods. The number of periods is set by the “periods
for averaging” variable. The length of the averaging period will be “seconds of
integration” X “number of periods averaged”.
INST MEDIUM INT TEMP
This parameter is the instantaneous temperature of the medium as measured by the
monitor during a single integration period.
AVE MEDIUM INT TEMP
This parameter is an average of the instantaneous temperature of the medium as
measured by the monitor during a number of integration periods. The number of
periods is set by the “periods for averaging” variable. The length of the averaging
period will be “seconds of integration” X “number of periods averaged”.
INSTANT SOUND SPEED
This parameter is the instantaneous speed of sound through the medium as measured
by the monitor during a single Integration Period.
AVG SOUND SPEED
This parameter is an average of the instantaneous speed of sound values as measured
by the monitor during a number of Integration Periods. The number of periods is set
by the “periods for averaging” variable. The length of the averaging period will be
“seconds of integration” X “number of periods averaged”.
AVG SPAN HIGH
This parameter is an average of the instantaneous span high values as measured by the
monitor during a number of Integration Periods. The number of periods is set by the
“periods for averaging” variable. The length of the averaging period will be “seconds
of integration” X “number of periods averaged”.
AVG SPAN LOW
This parameter is an average of the instantaneous span low values as measured by the
monitor during a number of Integration Periods. The number of periods is set by the
“periods for averaging” variable. The length of the averaging period will be “seconds
of integration” X “number of periods averaged”.
6-30
Summary of Contents for ULTRAFLOW 150
Page 2: ...DOCUMENT NO 1900 0100 01 REV E ...
Page 24: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank 4 6 ...
Page 75: ...SECTION 7 0 INSTALLATION This page intentionally left blank 7 7 ...
Page 76: ......
Page 85: ...SECTION 8 0 SYSTEM CALIBRATION AND ADJUSTMENT This page intentionally left blank 8 9 ...
Page 86: ......
Page 89: ...APPENDIX A SITE SPECIFICATION DATA SHEETS ...
Page 90: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank ...
Page 95: ...APPENDIX B MAINTENANCE CHECK SHEETS ...
Page 96: ... This page intentionally left blank ...
Page 101: ...APPENDIX C SPARE PARTS ...
Page 102: ... This page intentionally left blank ...
Page 105: ......
Page 106: ...APPENDIX D DRAWINGS ...
Page 107: ... This page intentionally left blank ...
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ...APPENDIX E ENHANCED SERIAL PORT COMMUNICATION PROTOCOL ...