SECTION 5.0, TRANSDUCER INTERFACE ENCLOSURE
logic. Bus transceivers are used to convert logic from one level to another
wherever needed.
5.2.3.6 Purge Switch Circuits
The signals from the purge switch enter the Flow Mother Board through the
Preamp Board connector. They are passed through an ESD/buffer network to the
computer module. A logic LOW indicates normal purge air flow.
5.2.3.7 On Board D.C. Supplies
The Flow Mother Board has circuits to produce the following D.C. supply
voltages: 2.5 VDC---Used for FPGA Core functions; 3.3 VDC---Used for FPGA
I/O functions; 5.0 VDC---Used for analog functions; -5.0 VDC---Used for analog
functions; 130 VDC---Used for short range transducer bias.
5.2.3.8 External Serial Data Connections
The Flow Mother Board supports both synchronous and asynchronous serial
devices. All synchronous serial devices are for use by Ultraflow 150 components,
most of which are on TIE circuit boards external to the Flow Mother Board.
The Flow Mother Board supports one RS232C asynchronous serial port for use
with Teledyne Monitor Labs’ PC to Modbus software. This port is accessible
through DB9 connector J9.
5.2.3.8.1 Synchronous Serial Devices
The Flow Mother Board employs 11 synchronous serial device channels. The
Serial Peripheral Interface (SPI) standard is used to convey data to the I/O
devices. The Computer Module Board acts as the master SPI device and hence
controls access. All of the devices are on external circuit boards with the
exception of the real time clock. The devices are listed below.
1)
Upstream Path A Preamp Gain
2)
Downstream Path A Preamp Gain
3)
Upstream Path B Preamp Gain (X Pattern only)
4)
Downstream Path B Preamp Gain (X Pattern only)
5)
Local User Interface LED Display (Optional)
6)
Local User Interface Keypad (Optional)
7)
Six Point I/O Analog Output DAC1 (Optional)
8)
Six Point I/O Analog Output DAC2 (Optional)
9)
Six Point I/O Isolator Inputs (Optional)
10)
Six Point I/O Relay Outputs (Optional)
11)
Real Time Clock (Internal to Flow Mother Board
)
5.2.3.8.2 TIE Asynchronous Serial Port
Flow Mother Board DB9 connector J9 is an RS232C serial port. This is a three
wire serial port that does not support frame control. A custom Modbus protocol is
5-7
Summary of Contents for ULTRAFLOW 150
Page 2: ...DOCUMENT NO 1900 0100 01 REV E ...
Page 24: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank 4 6 ...
Page 75: ...SECTION 7 0 INSTALLATION This page intentionally left blank 7 7 ...
Page 76: ......
Page 85: ...SECTION 8 0 SYSTEM CALIBRATION AND ADJUSTMENT This page intentionally left blank 8 9 ...
Page 86: ......
Page 89: ...APPENDIX A SITE SPECIFICATION DATA SHEETS ...
Page 90: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank ...
Page 95: ...APPENDIX B MAINTENANCE CHECK SHEETS ...
Page 96: ... This page intentionally left blank ...
Page 101: ...APPENDIX C SPARE PARTS ...
Page 102: ... This page intentionally left blank ...
Page 105: ......
Page 106: ...APPENDIX D DRAWINGS ...
Page 107: ... This page intentionally left blank ...
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ...APPENDIX E ENHANCED SERIAL PORT COMMUNICATION PROTOCOL ...