ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR
CAL CYCLE
Initiated by:
Digital Inputs,
Enhanced
Remote Panel,
Local User
Interface, or
System Clock
Initiates an internally timed calibration cycle consisting of the
three CALIBRATION modes and their supporting ACQUIRE
modes. The order is ZERO-ACQUIRE, ZERO, SPANLOW-
ACQUIRE, SPANLOW, SPANHIGH-ACQUIRE and
SPANHIGH. The combined length of each ACQUIRE/CAL
mode is set by the number of periods for ZERO, SPANLOW,
and SPANHIGH variables. If periods for ZERO are 10 and
periods for AVERAGE are 6, the ZERO ACQUIRE mode will
last for 6 integration periods and ZERO will last for 4 integration
periods. To defeat any of the CAL modes set the number of
periods for that mode equal to zero. See timing diagram.
The MIO mode follows the TIE mode. It
changes modes when it sees that the TIE
has been in the new mode for one full
period. The MIO mode will be one
integration period behind the TIE
throughout the entire CAL CYCLE. The
analog outputs will be updated at the end
of each integration period. The relays will
be updated at each MIO mode change.
See timing diagram.
TEST ZERO
SCALE
Initiated by:
Digital Inputs or
Enhanced Remote
Panel
The TIE ignores this mode
request. The TIE will remain in
whatever mode it was in.
This is
an MIO test only.
The MIO enters this mode on receipt of the request. This mode will drive
all analog outputs regardless of configuration to their ZERO scale. If
actuated the NORMAL relay will drop out to indicate that other than
NORMAL data is present on the outputs. The MIO will stay in this mode
until another mode request is issued.
TEST MID
SCALE
Initiated by:
Digital Inputs or
Enhanced
Remote Panel
The TIE ignores this mode
request. It will remain in
whatever mode it was in.
This is
an MIO test only.
The MIO enters this mode on receipt of the request. This mode will drive
all analog outputs regardless of configuration to their MID scale. If
actuated the NORMAL relay will drop out to indicate that other than
NORMAL data is present on the outputs. The MIO will stay in this mode
until another mode request is issued.
TEST FULL
SCALE
Initiated by:
Digital Inputs or
Enhanced
Remote Panel
The TIE ignores this mode
request. It will remain in
whatever mode it was in.
This is
an MIO test only.
The MIO enters this mode on receipt of the request. This mode will drive
all analog outputs regardless of configuration to their FULL scale. If
actuated the NORMAL relay will drop out to indicate that other than
NORMAL data is present on the outputs. The MIO will stay in this mode
until another mode request is issued.
Note: ACQUIRE modes are transitional modes designed to signal that the data from the monitor
is in transition from one mode to another. For instance, ZERO ACQUIRE indicates that
the monitor is producing data that is a mixture of ZERO and the previous mode.
ACQUIRE modes last for the averaging period. This time can be determined by the
following formula:
AVERAGING PERIOD = Integration Periods in Average * Integration Period
6.4.7 Output Module Configuration Menu
This sub-menu is password protected. It contains three screens for
configuring the analog and digital outputs of the Multi I/O Board. The
screens titled “ANALOG OUTPUTS 1,2” and “ANALOG OUTPUTS
3,4” configure their respective analog output parameter mapping and
scaling. The “RELAY ASSIGNMENT” screen configures the conditions
for closure of the eight output relays. Use the arrow keys to select
OUTPUT MODULE CONFIG then press “
↵
”
. This will prompt you to
choose one of the three screens. Make your selection with the arrow keys
and press enter.
6-16
Summary of Contents for ULTRAFLOW 150
Page 2: ...DOCUMENT NO 1900 0100 01 REV E ...
Page 24: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank 4 6 ...
Page 75: ...SECTION 7 0 INSTALLATION This page intentionally left blank 7 7 ...
Page 76: ......
Page 85: ...SECTION 8 0 SYSTEM CALIBRATION AND ADJUSTMENT This page intentionally left blank 8 9 ...
Page 86: ......
Page 89: ...APPENDIX A SITE SPECIFICATION DATA SHEETS ...
Page 90: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank ...
Page 95: ...APPENDIX B MAINTENANCE CHECK SHEETS ...
Page 96: ... This page intentionally left blank ...
Page 101: ...APPENDIX C SPARE PARTS ...
Page 102: ... This page intentionally left blank ...
Page 105: ......
Page 106: ...APPENDIX D DRAWINGS ...
Page 107: ... This page intentionally left blank ...
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ...APPENDIX E ENHANCED SERIAL PORT COMMUNICATION PROTOCOL ...